RF OUTPUT SPECIFICATIONS
Frequency | |
---|---|
Output Range | 140 kHz to 2.5 GHz |
Resolution | 0.1 Hz |
Phase [email protected] 1 GHz | <-90 dBc/Hz, 1 kHz Offset |
<-95 dBc/Hz, 10 kHz Offset | |
<-110 dBc/Hz, 100 kHz Offset | |
Noise Floor | -140 dBm/Hz @ 10 MHz Offset |
Internal Reference | +/- 0.2 ppm initial accuracy |
+/- 0.8 ppm/year aging | |
+/- 1.0 ppm temperature stability | |
Amplitude & Power | |
Output Range | -140 dBm to +10 dBm P.E.P. |
Resolution | 0.1 dB |
Accuracy | +/-1.0 dB (>-110 dBm), +/-2.0 dB (≤ -110 dBm) |
Gain Resolution, RF Playback | 0.25 dB |
Maximum IMD level | -60 dBm typ. (two -5 dBm tones) |
Harmonics NOTE : All specifications apply for √(12+Q2)=0.5FS |
-30 dBc @ 0 dBm |
Baseband | |
1 MHz to 20 MHz Real-Time BW | |
Sample Rate | 25 MS/s |
Dynamic Range | 80 dB SFDR |
Output Resolution | 16-bit |
CONNECTIVITY
RF Connectors | |
---|---|
RF Output (50Ω) | 1 x N Female |
+20 dBm, +/- 16V DC, AC Coupled | |
LO | |
LO Input | 1 x SMA Female |
Freq. 150 MHz / 2.5 GHz Level 0/+5 dBm, Max.: +12 dBm |
|
LO Output | 1 x SMA Female |
Freq. 150 MHz / 2.5 GHz Level 0/+5 dBm, Max.: +20 dBm |
|
10 MHz Reference | |
10 MHz REF Input (50Ω) | 1 x SMA Female |
Freq. 10 MHz Level 0/+10 dBm, Max.: +15 dBm |
|
10 MHz REF Output (50Ω) | 1 x SMA Female |
Freq. 10 MHz Level +2 dBm, Max.: +10 dBm |
|
100 MHz Sample Clock | |
Clock Input (50Ω) | 1 X SMA Female |
Freq. 100 MHz Level 0/+10 dBm, Max.: +12 dBm |
|
Clock Output (50Ω) | 1 X SMA Female |
Freq. 100 MHz Level +4 dBm, Max.: +20 dBm |
|
Trigger/Sync Input(s) & Output(s) | |
Input (50Ω) | 2 x SMA Female |
Freq.: PULSE Level: TTL 5 V TOL, Max.: -0.5/5.5 V |
|
Output (50Ω) | 1 x SMA Female |
Freq.: PULSE Level: TTL 5 V TOL, Max.: -0.5/5.5 V |
|
Ethernet | |
1 x 10/100/1000 Mbps RJ-45 LAN Port | |
Peripheral | |
4 x USB 2.0/1.1 Type A Peripheral Ports (Back) | |
2 x USB 2.0/1.1 Type A Peripheral Ports (Front) | |
Display | |
1 x DBHD-15F VGA Port | |
Compliance | |
FCC 47 part 15 Class A | |
European Directive 98/336/EEC Class A (Emissions) | |
European Directive 2002/95/EC (RoHS) |
ENVIRONMENTAL
Warm-Up Time | |
---|---|
60 min @ 20˚C (typ.) | |
Weight | |
15 kg (33.06 lbs) | |
Size (H x W x D) | |
Unit | 2 U |
Rackmount | 48 cm (19 in) |
Deep | 40 cm (16 in) |
Removable rack ears and handles | |
Temperature | |
Operating | 5˚C (41˚F) to 45˚C (113˚F) |
Storage | -20˚C (-4˚F) to 70˚C (158˚F) |
Relative Humidity | |
10% to 90% (non-condensing) | |
Calibration | |
2 years | |
Warranty | |
24 months | |
Power | |
90-264 VAC | |
50/60 Hz | |
125 Watts (typ.) | |
IEC 60320-C14 Power connector inlet | |
IEC 60320-C13 to NEMA 5-15P, 3 m (9.8 ft) North American power cord included | |
IEC 60320-C13 to CEE 7/7, 3 m (9.8 ft) European power cord included |